Qubit Protocol
Based on Putnam Lab Protocol written by M. Schedl: Qubit
Broad Range dsDNA, Broad Range RNA, and High Sensitivity RNA
All Steps are the same whichever quantification method you are using
- Determine n number: Number of samples + 2 + % error
- Create master mix of Qubit Buffer and Quant-IT Reagent
- n μl of Quant-IT Reagent
- 199 * n μl of Buffer
- Vortex and spin down master mix
- Set up thin walled tubes: enough for each of your samples plus 2 more for standards
- Add 190μl of master mix to the 2 tubes for the standards
- Add 199μl of master mix to each of your sample tubes
- Take standards out of the fridge (they say HP for the Putnam Lab) and add 10μl of Standard 1 to the first standard thin walled tube and add 10μl of Standard 2 to the second standard thin walled tube.
- Add 1μl of sample to the corresponding thin walled sample tube.
- Vortex and spin down all tubes.
- Choose appropriate assay in the Qubit and read standard tubes.
- If your standard values are not near the range of values in these graphs then your master mix was probably made wrong.
- Read each sample tube one at a time. Our settings are 1μl of sample and we want the output to be ng/μl.
- Read your samples twice.
- All components are safe and can be disposed of in the tip buckets or trash.
Holobiont Integration project
Master Mix Calculations:
4 coral fragments (4 soft homogenized, 4 hard homogenized = 8 total):
- 8 samples + 2 standards + 0.2% error = 10.2 µl Quant-IT Reagent
- 199 * 10.2 = 2,029.8 µl Qubit buffer
8 coral fragments (8 soft homogenized, 8 hard homogenized = 16 total):
- 16 samples + 2 standards + 0.2% error = 18.2 µl Quant-IT Reagent
- 199 * 18.2 = 3,621.8 µl Qubit buffer
10 coral fragments (10 soft homogenized, 10 hard homogenized = 20 total):
- 20 samples + 2 standards + 0.2% error = 22.2 µl Quant-IT Reagent
- 199 * 22.2 = 4,417.8 µl Qubit buffer
Written on May 31, 2019